MINIVOL LVR V/L Ratio Tester


The Ultimate Volatility Tester for GasolineProduct | Small | MINIVOL LVR

MINIVOL LVR is a tester for the automatic determination of the vapor-liquid ratio temperature. Non-viscous liquids like gasoline, solvents and other highly volatile compounds can be tested with a high precision range of 20 to 80°C (68 to 176°F). The normal pressure for the test can be adjusted between 50 kPa and 200 kPa.

NEW DESIGN
MINIVOL LVR vapor pressure tester was designed for the automatic determination of the Vapor/Liquid-Ratio temperature. Non-viscous liquids like gasoline, solvents and other highly volatile compounds can be tested with high precision in a range of 20 to 80°C (68 to 176°F). The nominal pressure for the test can be adjusted between 50 kPa and 200 kPa (7.25 and 29.01 psi).

FULLY AUTOMATIC
The instrument performs fully automatically with a sample volume of only 5 mL. The syringe is filled with the sample and inserted into the Luer-lock on the right hand side of the instrument.

SIGNIFICANCE AND USE (ASTM D5188)
The tendency of fuel to vaporize in automotive engine fuel systems is indicated by the Vapor/Liquid-Ratio of the fuel. Automotive fuel specifications generally include T (V/L=20) limits to ensure products of suitable volatility performance. Studies of vapor lock behavior are best performed by measuring multiple point measurements at different nominal pressures.

TEST METHOD
The measuring chamber with a volume of 15 mL is evacuated and heated above the expected temperature. After flushing the chamber with sample and further evacuation, the exact amount of liquid for the first point is injected and the temperature is decreased until nominal pressure (e.g. 101.3 kPa or 14.69 psi) is observed. The results are calculated according to
V/L=(V - v)/v

V... Volume of the test chamber
v... injected sample volume

MULTIPLE POINT MEASUREMENT
For multiple point evaluation, the test is started with V/L=100 and the next steps of V/L=80 down to V/L=4 are measured consecutively by adding the respective amount of sample.

MAGNETIC STIRRER
A magnetic stirrer inside the measuring chamber serves for fast equilibrium and short measuring time.