Grabner/Reid Vapor Pressure On-Line Analyzer
Features
- On-line Compliance Monitoring for Vapor Pressure of Gasoline, LPG and Crude Oil and Vapor/Liquid-Ratio Temperature of Gasoline (V/L=20)
- One Single Instrument for up to 2 Sample Streams
- Multiple Point Measurements at Different Nominal Pressures
- Automatic Calibration & Offset Correction
- Temperature Controlled Sample Conditioning System
- 7 Minutes Cycle Time
- Explosion Proof-Class I
- NFPA 496 (US)
- EN 60079-2 and EN 60079-0 (EU)
VAPOR PRESSURE TESTER STANDARDS
- ASTM D6378 - Vapor Pressure of Gasoline (VP4)
- ASTM D6377 - Vapor Pressure of Crude Oil
- ASTM D6897 - Latest standard for Vapor Pressure of LPG
- ASTM D5191
- ASTM D5188 - V/L-Ratio: T (V/L = 20)
- ASTM D4953
- ASTM D1267
- ASTM D323
- EN 13016 - 1+2
- IP 394, 409 & 481
- JIS K2258-2, SHT 0769, GOST 52340
|
Technical Data
Temperature Range
|
20 to 60°C (68 to 140°F) |
| Pressure Range |
0 to 1000 kPa (0 to 145 psi) |
| Power Requirements |
100/120/230/240 V AC, 50/60 Hz, 65 W |
| Field Application |
12 V/4A DC (vehicle battery) |
| Physical Dimensions |
580 x 1060 x 260 mm (23.7" x 43.3" x 10.6") |
| Weight |
50 kg (110 lb) |
| Precision Data |
MINIVAP 37.8°C @ 70 kPa) |
ASTM D323 (Wet Reid Bomb) |
| Repeatability |
±0.3 kPa (0.04 psi) |
±3.2 kPa (0.46 psi) |
| Reproducibility |
±0.7 kPa (0.10 psi) |
±5.2 kPa (0.75 psi) |
Principle
The sample flows from the sample conditioning system via a quick connector (3) and sample inlet valve (4) into the measuring chamber. The automatic sample introduction and volume adjustments are made by a piston with an integrated pressure transducer (1). The measuring chamber (2) with a total volume of 5 mL is rinsed with 2.5 mL and filled with 1 mL. After closing the valve (4), the single or triple expansion to 5 mL (with the sample under vacuum condition) is obtained by further strokes of the piston. The temperature of the measuring cell is controlled with a high-power thermoelectric module (5) and measured with a precision Platinum sensor (6). To achieve fast pressure equilibrium, especially for crude oil, the measuring unit is rocked during the test procedure.
Automatic piston lubrication is processed with every piston stroke.

Continuous product development may make it necessary to change product specifications without notice. |